Users Manual Part 2

379Initial configuration
Roche Diagnostics
cobas
®
pulse · Software version 01.03 · User Assistance · Publication version 1.0
Dependencies
Configuring Shift length (in hours) is only necessary
when the QC algorithm configuration item value is Shift
or Shift (alternating QC level).
Configuration
You configure the Shift length (in hours) configuration
item using the following:
Data management system (e.g., cobas
®
infinity POC)
Barcode (generated via cobas
®
infinity edge)
u Related topics
QC algorithm (376)
Shift start (379)
Shift start
This configuration item specifies the start time of a shift.
This is required when a QC lockout is based on the
changeover of a working shift.
Shift start works in conjunction with the Shift length
(in hours) configuration item.
Item ID
QC_SHIFT_START
Available values/format
0 - 23 hours (default: 6 hours)
Values must be integers.
Dependencies
Configuring Shift start is only relevant when the QC
algorithm configuration item value is Shift or Shift
(alternating QC level).
Configuration
You configure the Shift start configuration item using
the following:
Data management system (e.g., cobas
®
infinity POC)
Barcode (generated via cobas
®
infinity edge)
u Related topics
QC algorithm (376)
Shift length (in hours) (378)
10 Initial configuration