• Time of day (380)
QC interval (in hours)
This configuration item specifies the time period (hours)
for which a positive QC result is valid. A QC lockout is
enabled after this period is exceeded.
QC interval (in hours) works in conjunction with the
QC algorithm configuration item.
Item ID
QC_HOURS
Available values/format
1-9999 hours (default: 1000 hours).
Values must be integers.
Dependencies
Configuring QC interval (in hours) is only necessary
when the QC algorithm configuration item value is
Time interval or Time interval (alternating QC level).
Configuration
You configure QC interval (in hours) configuration
item using the following:
• Data management system (e.g., cobas
®
infinity POC)
• Barcode (generated via cobas
®
infinity edge)
u Related topics
• QC algorithm (376)
Shift length (in hours)
This configuration item specifies the length of a working
shift in hours. This is required when a QC lockout is
based on the changeover of a working shift.
Shift length (in hours) works in conjunction with the
Shift start configuration item.
Item ID
QC_SHIFT_LENGTH
Available values/format
The value must be an even divisor of 24:
1, 2, 3, 4, 6, 8 (default), 12, 24 hours.
Values must be integers.