Users Manual Part 2
377Initial configuration
Roche Diagnostics
cobas
®
pulse · Software version 01.03 · User Assistance · Publication version 1.0
QC lockout is enabled when the last QC test
performed is out of range. Consequently if no
successful QC test has been performed, the
instrument will be in QC lockout.
• Shift
[SHIFT]
A QC lockout is activated with each change of shift. It
works in conjunction with the Shift length (in
hours) and Shift start configuration items.
• Shift (alternating QC level)
[SHIFT_ROTATE]
The same as Shift, but the QC test uses the level 1
and the level 2 QC materials alternately rather than
both levels.
• Test strip count
[STRIP_COUNT]
QC lockout is enabled after a predetermined number
of successful tests have been performed (excluding
QC tests). It works in conjunction with the Test strip
count configuration item.
• Test strip count (alternating QC level)
[STRIP_COUNT_ROTATE]
The same as Test strip count, but the QC test uses
level 1 and level 2 QC materials alternately rather than
both levels.
• Time of day
[TIME_OF_DAY]
QC lockouts are enabled at specified times of day (up
to 6). It works in conjunction with the Time of day
configuration item.
• Time of day (alternating QC level)
[TIME_OF_DAY_ROTATE]
The same as Time of day, but the QC uses L1 and L2
controls alternately rather than both controls.
Dependencies
If Country = "US", the value of QC algorithm cannot be
set to Disabled.
Configuration
You configure the QC algorithm configuration item
using the following:
• Data management system (e.g., cobas
®
infinity POC)
• Barcode (generated via cobas
®
infinity edge)
u Related topics
• QC interval (in hours) (378)
• Shift length (in hours) (378)
• Shift start (379)
• Test strip count (380)
10 Initial configuration