Users Manual Part 2

376 QC configuration
Roche Diagnostics
cobas
®
pulse · Software version 01.03 · User Assistance · Publication version 1.0
QC configuration
In this section
QC algorithm(376)
Show numeric QC result value(382)
QC range handling(382)
QC algorithm
In this section
QC algorithm(376)
QC interval (in hours)(378)
Shift length (in hours)(378)
Shift start(379)
Test strip count(380)
Time of day(380)
QC tests required for new test strip lots(381)
QC algorithm
This configuration item specifies how QC lockout is
enabled on the instrument.
Item ID
QC_ALGORITHM
Available values/format
Disabled
[ALWAYS_OK]
QC lockout is disabled.
Time interval
[HOURS]
QC lockout is enabled after a specified number of
hours since the last successful QC test. was
performed. It works in conjunction with the QC
interval (in hours) configuration item.
Time interval (alternating QC level)
[HOURS_ROTATE]
The same as Time interval, but the QC test uses the
level 1 and the level 2 QC materials alternately rather
than both levels.
Until failed QC (Default)
[LAST_QC_OK]
10 Initial configuration