Users Manual Part 2
376 QC configuration
Roche Diagnostics
cobas
®
pulse · Software version 01.03 · User Assistance · Publication version 1.0
QC configuration
In this section
QC algorithm(376)
Show numeric QC result value(382)
QC range handling(382)
QC algorithm
In this section
QC algorithm(376)
QC interval (in hours)(378)
Shift length (in hours)(378)
Shift start(379)
Test strip count(380)
Time of day(380)
QC tests required for new test strip lots(381)
QC algorithm
This configuration item specifies how QC lockout is
enabled on the instrument.
Item ID
QC_ALGORITHM
Available values/format
• Disabled
[ALWAYS_OK]
QC lockout is disabled.
• Time interval
[HOURS]
QC lockout is enabled after a specified number of
hours since the last successful QC test. was
performed. It works in conjunction with the QC
interval (in hours) configuration item.
• Time interval (alternating QC level)
[HOURS_ROTATE]
The same as Time interval, but the QC test uses the
level 1 and the level 2 QC materials alternately rather
than both levels.
• Until failed QC (Default)
[LAST_QC_OK]
10 Initial configuration