354 Level 2 errors
Roche Diagnostics
cobas
®
pulse · Software version 01.03 · User Assistance · Publication version 1.0
Solution
Ensure that the instrument is connected to the network
using the Network status test option, and that the
connection is stable. Restart the software update via
cobas
®
infinity edge.
Related L1 message
I-AC
u About level 1 messages and level 2 errors (209)
US-0017: Measuring unit software update error
Problem
The low-level measuring unit firmware update failed.
Solution
Contact your Roche representative.
Related L1 message
I-AC
u About level 1 messages and level 2 errors (209)
US-0018: Software update failed
Problem
The software update package was rejected by the
instrument due to a signature verification error.
Solution
Contact your Roche representative.
Related L1 message
I-AC
u About level 1 messages and level 2 errors (209)
US-0019: Operating system update failed
Problem
An operating system software update failed.
Solution
Ensure that the instrument is connected to the network
using the Network status test option, and that the
connection is stable. Restart the software update via
cobas
®
infinity edge.