Users Manual Part 2
352 Level 2 errors
Roche Diagnostics
cobas
®
pulse · Software version 01.03 · User Assistance · Publication version 1.0
US-0010: Measuring unit error
Problem
A measuring unit software update error occurred.
Solution
Restart the instrument.
Ensure that the instrument is connected to the network
using the Network status test option, and that the
connection is stable.
Restart the software update via cobas
®
infinity edge.
Related L1 message
G-AA
u About level 1 messages and level 2 errors (209)
US-0011: Measuring unit error
Problem
A measuring unit software update error occurred.
Solution
Restart the instrument.
Ensure that the instrument is connected to the network
using the Network status test option, and that the
connection is stable.
Restart the software update via cobas
®
infinity edge.
Related L1 message
G-AA
u About level 1 messages and level 2 errors (209)
US-0012: High-level measuring unit firmware update failed
Problem
The measuring unit firmware update could not be
completed via the intended high-level software interface.
Solution
No further action required since the update was
performed using a low-level software interface.
Related L1 message
I-AC
9 Error messages