Users Manual Part 2
346 Level 2 errors
Roche Diagnostics
cobas
®
pulse · Software version 01.03 · User Assistance · Publication version 1.0
If this problem persists while testing other patients,
contact your Roche representative.
Related L1 message
G-AC
u About level 1 messages and level 2 errors (209)
MU-0979: Removal of test strip during test
Problem
The measuring unit detected a signal range violation.
Solution
The test strip was removed during the test.
Repeat the test with a new test strip. Do not remove test
strip during processing.
Related L1 message
D-AL
u About level 1 messages and level 2 errors (209)
MU-0980: Signal range violation
Problem
The measuring unit detected a signal range violation.
Solution
Repeat the test with a new test strip.
If this problem persists while testing other patients,
contact your Roche representative.
Related L1 message
G-AC
u About level 1 messages and level 2 errors (209)
MU-0981: Signal range violation
Problem
The measuring unit detected a signal range violation.
Solution
Repeat the test with a new test strip.
9 Error messages