Users Manual Part 2

343Error messages
Roche Diagnostics
cobas
®
pulse · Software version 01.03 · User Assistance · Publication version 1.0
Solution
Repeat the test with a strip from another container.
If this does not solve the issue, repeat the test with
another strip and another instrument.
Related L1 message
D-AF
u About level 1 messages and level 2 errors (209)
MU-0962: Test strip error
Problem
A test strip error occurred.
Solution
Repeat the test with a new test strip. If this problem
persists while testing other patients, contact your Roche
representative.
Related L1 message
D-AF
u About level 1 messages and level 2 errors (209)
MU-0963: Test strip error
Problem
A test strip error occurred.
Solution
Repeat the test with a strip from another container.
If this does not solve the issue, repeat the test with
another strip and another instrument.
Related L1 message
D-AF
u About level 1 messages and level 2 errors (209)
MU-0971: Signal range violation
Problem
The measuring unit detected a signal range violation.
Solution
Repeat the test with a new test strip.
9 Error messages