Users Manual Part 2

342 Level 2 errors
Roche Diagnostics
cobas
®
pulse · Software version 01.03 · User Assistance · Publication version 1.0
Apply sufficient sample when requested without moving
the test strip. If this problem persists, contact your Roche
representative.
Related L1 message
S-AB
u About level 1 messages and level 2 errors (209)
MU-0952: Sample application error
Problem
The instrument detected an unsteady sample application,
or the test strip was moved during sample application or
sample detection.
Solution
Repeat the test with a new test strip.
Apply sufficient sample when requested without moving
the test strip. If this problem persists, contact your Roche
representative.
Related L1 message
S-AB
u About level 1 messages and level 2 errors (209)
MU-0960: Test strip error
Problem
A test strip error occurred.
Solution
Repeat the test with a new test strip. If this problem
persists while testing other patients, contact your Roche
representative.
Related L1 message
D-AA
u About level 1 messages and level 2 errors (209)
MU-0961: Test strip error
Problem
A test strip error occurred.
9 Error messages