Users Manual Part 2
341Error messages
Roche Diagnostics
cobas
®
pulse · Software version 01.03 · User Assistance · Publication version 1.0
MU-0945: Sample application error
Problem
The instrument detected an unsteady sample application,
or the test strip was moved during sample application or
sample detection.
Solution
Repeat the test with a new test strip.
Apply sufficient sample when requested without moving
the test strip. If this problem persists, contact your Roche
representative.
Related L1 message
S-AB
u About level 1 messages and level 2 errors (209)
MU-0947: Sample application error
Problem
The instrument detected an unsteady sample application,
or the test strip was moved during sample application or
sample detection.
Solution
Repeat the test with a new test strip.
Apply sufficient sample when requested without moving
the test strip. If this problem persists, contact your Roche
representative.
Related L1 message
S-AB
u About level 1 messages and level 2 errors (209)
MU-0951: Sample application error
Problem
The instrument detected an unsteady sample application,
or the test strip was moved during sample application or
sample detection.
Solution
Repeat the test with a new test strip.
9 Error messages