MU-0930: Sample application error
Problem
The sample was applied to early.
Solution
Repeat the test with a new test strip.Apply sufficient
sample when requested without moving the test strip.
If this problem persists, contact your Roche
representative.
Related L1 message
S-AB
u About level 1 messages and level 2 errors (209)
MU-0940: Sample application error
Problem
The instrument could not detect a sample application
within the given timeout.
Solution
Repeat the test with a new test strip.Apply sufficient
sample when requested without moving the test strip. If
this problem persists, contact your Roche representative.
Related L1 message
S-AB
u About level 1 messages and level 2 errors (209)
MU-0941: Sample application error
Problem
The instrument detected an unsteady sample application
or insufficient sample volume was applied.
Solution
Repeat the test with a new test strip.
Apply sufficient sample when requested without moving
the test strip. If this problem persists, contact your Roche
representative.
Related L1 message
S-AB
u About level 1 messages and level 2 errors (209)