Users Manual Part 2

337Error messages
Roche Diagnostics
cobas
®
pulse · Software version 01.03 · User Assistance · Publication version 1.0
MU-0561: Measuring unit fail-safe activated
Problem
The measuring unit triggered a fail-safe logic for the
applied sample. The test failed.
Solution
Repeat the test with a new test strip. Follow proper usage
in the handling recommendations.
Related L1 message
S-AF
u About level 1 messages and level 2 errors (209)
MU-0562: Measuring unit fail-safe activated
Problem
The measuring unit triggered a fail-safe logic for the
applied sample. The test failed.
Solution
Repeat the test with a new test strip. Follow proper usage
in the handling recommendations.
Related L1 message
S-AF
u About level 1 messages and level 2 errors (209)
MU-0563: Measuring unit fail-safe activated
Problem
The measuring unit triggered a fail-safe logic for the
applied sample. The test failed.
Solution
Repeat the test with a new test strip. Follow proper usage
in the handling recommendations.
Related L1 message
S-AF
u About level 1 messages and level 2 errors (209)
9 Error messages