Users Manual Part 2
320 Level 2 errors
Roche Diagnostics
cobas
®
pulse · Software version 01.03 · User Assistance · Publication version 1.0
Solution
Restart the instrument and try again.
If the problem persists, contact your Roche
representative.
Related L1 message
G-AA
u About level 1 messages and level 2 errors (209)
MU-0341: Test strip ID decoding error
Problem
The instrument could not read the test strip ID (e.g. due
to dirt or scratches).
Solution
Eject the test strip and insert a new test strip. If the
problem persists, contact your Roche representative.
Related L1 message
D-AA
u About level 1 messages and level 2 errors (209)
MU-0342: Test strip ID decoding error
Problem
The instrument could not read the test strip ID (e.g. due
to dirt or scratches).
Solution
Eject the test strip and insert a new test strip. If the
problem persists, contact your Roche representative.
Related L1 message
D-AA
u About level 1 messages and level 2 errors (209)
MU-0343: Test strip insertion error
Problem
The measuring unit detected that a test strip was not
fully inserted or inserted upside down.
Solution
Eject the test strip and insert it correctly.
9 Error messages