Users Manual Part 2

302 Level 2 errors
Roche Diagnostics
cobas
®
pulse · Software version 01.03 · User Assistance · Publication version 1.0
Solution
Use the instrument at a higher temperature. The
temperature lockout will be deactivated as soon as the
instrument detects a supported instrument temperature.
Related L1 message
E-AA
u About level 1 messages and level 2 errors (209)
MU-0157: Temperature too high
Problem
The instrument has detected that the instrument
temperature is too high for testing and has activated the
temperature lockout.
Solution
Use the instrument at a lower temperature. The
temperature lockout will be deactivated as soon as the
instrument detects a supported instrument temperature.
Related L1 message
E-AB
u About level 1 messages and level 2 errors (209)
MU-0158: Measuring unit error
Problem
A measuring unit error occurred.
Solution
Repeat the test with a new test strip.
If the problem persists, contact your Roche
representative.
Related L1 message
G-AC
u About level 1 messages and level 2 errors (209)
MU-0159: Measuring unit error
Problem
A measuring unit error occurred.
9 Error messages