The barcode samples shown here are for illustration
purposes only. If printed out, they can be used to check
the barcode camera. However, they are not meant to be
used as a reference for size or resolution of real patient
or user ID barcodes.
When creating patient or user barcodes, always refer to
the relevant standard ISO/IEC 15416 and 15415 for size
and resolution requirements and to the specifications
listed in the table.
Recommended specification Comment
Print resolution 300 dpi preferred
200 dpi minimum
At 200 dpi issues with the wide-to-
narrow ratio may exist.
Reflective contrast 70% or greater Matte finish is preferred over glossfinish.
Symbol grade Grade C or above
Grade B is preferred
Symbol grades are A-F based on analysis
of several quality elements.
Depending on quality grade parameters
for a specific barcode, grade C may not
be sufficient when motion, reflection, or
poor lighting occur.
Module width
(minimum)
0.16 mm (1D barcodes)
0.20 mm (2D barcodes)
y Recommended specifications for barcode labels
i
Ensure that your printer can print the module
width of the barcode with the required resolution,
that is, that there is no mismatch between print
resolution and module width.
Supported barcode types
Codabar
Code 39
Code 93
Code 128
EAN 13
GS1 DataBar Limited
Interleaved 2 of 5 without
checksum