Users Manual Part 1

123Routine operation
Roche Diagnostics
cobas
®
pulse · Software version 01.03 · User Assistance · Publication version 1.0
c
Intervals between running QC tests are determined by
your healthcare facility.
d
m Glucose QC material
m Glucose test strips
m Waste container
m New pair of disposable gloves
j
m The instrument is turned on.
m You are logged on to the instrument.
u Logging on to the Glucose app (99)
r To perform a QC test
Tests
QC
1 On the Tests screen, tap the QC button.
QC test
Level 1 (low)
Level 2 (high)
Required
Required
2 On the QC test screen, do one of the following:
To perform a level 1 QC test, tap the Level 1
(low) button.
To perform a level 2 test, tap the Level 2 (high)
button.
I
When a QC test is required (due to a QC lockout),
the levels that you must test are indicated by
"Required".
3 To select the QC lot that you are using, do one of the
following:
In the Scan QC level 1 barcode screen / Scan
QC level 2 barcode screen, tap the tab and
then scan the QC lot number barcode.
On the Choose QC lot screen. tap the tab,
and then tap the lot number that you are using.
I
The tab might be deactivated in your
healthcare facility.
5 Routine operation