Performing a glucose test
One of the main functions of the instrument is to perform
glucose tests.
The procedure outlined in this task contains the main
steps that you perform when performing a glucose test.
However, depending on your instrument’s configuration,
there may be additional steps in the glucose test:
Patient ID confirmation (after entering the patient).
Sample type selection (after entering the patient /
after Patient ID validation).
Test strip ejection before confirming the test result
(not after).
Comment entry (before result confirmation).
Running another glucose test for the same patient
(after glucose test completion).
Entering insulin information (after glucose test
completion).
i
A colored test strip port light provides guidance
of instrument status and actions you must take.
In the case of a QC lockout, glucose testing cannot be
performed until the lockout has been removed.
Depending on how QC lockout is configured, you remove
it by performing and passing a QC test at level 1, level 2,
or level 1 and level 2.
i
If your instrument is configured to allow you to
override a QC lockout, you can perform a test
despite a QC lockout, when the QC lockout was
the result of a QC algorithm configuration
condition.
Your instrument can be configured to allow you
to override a QC lockout up to 9 times before you
must perform a QC test.
However, if the QC lockout was the result of a
failed QC test, the QC lockout cannot be
overridden.