Specifications
© 2001 Robinair, SPX Corporation
28
CONFIRMING THE SCALE CHECKLIST
Blank Display or No Warnings.
Check wire connections from scale assembly to the circuit board for configuration and
continuity.
Verify that the circuit board is receiving proper voltage (refer to the decal on the back of
the unit) and the two amp or 1/2 amp fuse is not blown. Use properly grounded active
electrical outlets only. Replace the circuit board or fuses as needed.
Total Amounts Not Accurate
Verify that nothing is touching, blocking, or interfering in any way with the scale assem-
bly, tank, or hoses. Confirm that the total weight on the scale assembly is not exceeding
80 lb/37 kg. If it is, reduce the amount of refrigerant in the tank.
Full (Full Tank Setting)
The compressor will shut off at the weight previously set. To check the weight of the
tank, scale, and refrigerant being recovered, follow the appropriate sections found under
USING THE DIAGNOSTIC MODE.
CORRECT WEIGHT VERIFICATION
1. Turn on the MAIN POWER switch.
2. Press SHIFT/RESET until "PROGRAM" displays.
NOTE: If the unit is just being turned on, the screen will default to display the selected
refrigerant type. Press SHIFT/RESET until "PROGRAM" displays.
3. When "PROGRAM" message displays, press and hold SHIFT/RESET, then press
ENTER. The "FUNC" message will display when the manual diagnostic mode has
been accessed properly.
4. Remove all weight from the scale platform.
5. Press 6 to turn the scale into a direct-reading weight scale.
NOTE: Any weight on the scale when 6 is pressed will not be shown on the display.
The unit automatically zeros the weight on the scale when 6 is pressed. If you remove the
weight, the display will show the change in total weight but will not show a negative sign.
6. Place a known weight of between 30-60 lbs./13-28 kg. on the scale. The display
will show the known weight "± .04lb./.02 kg."
7. If the scale does not read the weight accurately, recalibrate the scale and UL circuit
using the following instructions.
Maintenance Procedures
Important:
Check the
scale ac-
curacy
every
three
months.