User Manual

14
ERROR CODES AND CORRECTIVE ACTIONS
The instrument is supplied with self-diagnostic software to aid in
determining problems. The software will supply error messages that
direct the user toward specic corrective actions in the event of system
failure. Should corrective measures not correct the problem, the user is
advised to contact Robinair at 1-800-822-5561 for assistance.
SAMPLE FAULT: Signies unstable infrared readings due to
inconsistent sample gas supply or interference from strong electrical
elds.
Corrective Actions:
1. If the error code occurs during a sample cycle, verify the sample
hose is receiving sample gas from the source connection of at least
5 psig. Verify the instrument sample exhaust port and air intake
port are not clogged or restricted.
2. If the error code occurs during a calibration cycle, verify the air
intake port and sample exhaust port of the instrument are not
clogged or obstructed.
3. Keep the unit away from sources of strong electrical elds, such as
large compressors, running vehicles, etc. Move the instrument 3–5
feet away from the electrical eld source, and try again.
4. If the instrument is being used outside of the operating temperature
range of 39–122°F (4–50°C), allow the instrument to warm or cool
to the operating temperature range, and try again. The instrument
may require additional warm-up time after “power up” before use.
5. To clear the error code, power down the instrument. The error code
automatically clears.