Range Resoluon Accuracy
0Ω~10kΩ 1Ω ± (0.9% + 2d)
Resistor / Connuity
Range Resoluon Accuracy
0.4V~0.8V 0.1 V ± (1.0% + 3d)
Diode
Overload Protecon: AC 1000Vrms, 1000Vdc.
AC Current
Adjacent Conductor Inuence: < 0.08A/A.
Frequency Response: 50 ~ 60Hz (Sine Wave)
VoltSense: 80V ~ 1000V AC with 50Hz ~ 60Hz.
Connuity: Built-in buzzer sounds when measured resistance is less than 25Ω and sounds
o when measured resistance is more than 400Ω, Between 25Ω to 400Ω the buzzer maybe
sound or o either.
Overload Protecon: AC 1000Vrms, 1000Vdc.
Maximum Open Voltage: 1.6V.
Range Resoluon Accuracy
1.5A~200.0A 0.1 A ± (3.0% + 5d)
The proprietary informaon in this manual is protected by copyrights. Any photocopies,
reproducons or translaon to another language are not allowed unless it is permied.
And all rights are reserved.
The informaon in this manual is corre ct when prinng. However, RND will connuously
improve products and reserve the rights to change specicaons, equipment, and
maintenance procedures at any me without noce.