User Manual
10
Select Test Frequency
Open/Short Calibration
The default test frequency is 1kHz when the meter is turned on.
Press the Cal/Hz buon to select the test frequency.
In order to achieve the best measuring result, the calibraon has to be done
before measuring the DUT. To calibrate the meter, press and hold the Cal/Hz buon > 2 sec.
When “OPEn” appears on the display, make the test probes open, and press the Cal/
Hz buon to start open calibraon. During open calibraon, the “OPEn” blinks on
the display. About 15 seconds later, the result of the open calibraon appears on the
display. If the result shows "PASS", press the Cal/Hz buon to next step. If the result
shows "FAIL", press the Cal/Hz buon to exit calibraon mode.