User Manual
16
General Specications
Inductance
D & Q
θ
Denion: Q = 1/D = tanθ
Range: 2.000 to 2000
Minimum Resoluon: 0.001
Accuracy: ± (0.5% + 5) x (1+D), when D < 1 or Q > 1
Definition: θ = tan
-1
Q
Range: -90.0° to 90.0°
Minimum Resolution: 0.1°
Accuracy: ± (0.5% + 5)
Range 100Hz 1kHz 10kHz
120Hz
20.000uH N/A N/A N/A
200.00uH N/A N/A 0.5% + 30
[1]
2000.0uH N/A 0.5% + 30
[1]
0.5% + 5
20.000mH 0.5% + 30
[1]
0.5% + 5 0.5% + 5
200.00mH 0.5% + 5 0.5% + 5 0.5% + 5
2000.0mH 0.5% + 5 0.5% + 5 1.0% + 5
[2]
[1] Accuracy is specied aer subtract of the oset inductance.
[2] < 50dgt rolling.
[3] If D > 0.1, the accuracy should be mulplied by 1 + D
2