User Manual

34 35
EN EN
500μA
5000μA
50mA
500mA
10A
0.01μA
0.1μA
0.001mA
0.01mA
0.001A
50/60Hz
±(0.8% + 5 digits)
<5kHz
±(3% + 5 digits)
10A: 30 sec max with reduced accuracy.
All AC current ranges are specied from 5% of range to 100% of range.
5nF (3) 0.001nF
50nF 0.01nF
500nF 0.1nF
5μF 0.001μF
50μF 0.01μF
500μF 0.1μF
10mF 0.01mF
±(1.5% + 20 digits)
±(1.5% + 8 digits)
±(1.0% + 8 digits)
±(1.5% + 15 digits)
±(1.0% + 8 digits)
±(1.5% + 8 digits)
±(2.5% + 20 digits)
(3): With a lm capacitor or beer, using relave mode (REL ) to zero residual.
50Hz
500Hz
5kHz
50kHz
500kHz
5MHz
10MHz
0.001Hz
0.01Hz
0.0001kHz
0.001kHz
0.01kHz
0.0001MHz
0.001MHz
±(0.01% + 5 digits)
Unspecied
Sensivity: 2V rms min. at 20% to 80% duty cycle and <100kHz; 5Vrms min at 20% to 80% duty
cycle and >100kHz.
10.00-10kHz 0.01-0.001kHz ±(0.5% reading)
Sensivity: 2Vrms.
0.1 to 99.90%
0.01%
±(1.2% + 5 digits)
Pulse width: 100μs-100ms, Frequency: 5Hz to 150kHz.
-200 to 1350°C
-328 to 2462°F
0.1°C
0.1°F
±(1.0% + 3.0°C)
±(1.0% + 5.4°C)
Does not include error of the thermocouple probe.
Accuracy specicaon assumes ambient temperature stable to ±1°C.
Use a long me, reading will increase 2°C.
<-50°C Temp Rang accuracy ±(3%+5°C).
10-2.General Specicaons
Enclosures Double molded
Shock 6.5 feet (2 meters)
Waterproof IP67
Diode Test Test current of 1.5mA maximum, open circuit voltage 3.2V
DC typical.
Connuity Check Audible signal will sound if the resistance is less than 25Ω(ap
prox.), test current <0.35mA.
PEAK Captures peaks >1ms
Temperature Sensor Requires type K thermocouple
Input Impendance >10MΩ VDC & >9MΩ VAC
AC Response True rms
AC Response The term stands for “Root-Mean-Square,” which represents the
method of calculaon of the voltage or current value.
Average responding mulmeters are calibrated to read correctly
only on sine waves and they will read inaccurately on non-sine
wave or distorted signals. True rms meters read accurately
on either type of signal.
ACV Bandwidth 50Hz to 20000Hz
Display 50,000 Counts TFT LCD
Overrange Indicaon “OL” is displayed
Auto Power O 5-30minutes (approximately) with disable feature
Polarity Automac (no indicaon for posive); Minus (-) sign for negave
Measurement Rate 20 mes per second, nominal
Low Baery Indicaon “ ” is displayed if baery voltage drops below operang voltage
Baery One 7.4 volt (NEDA 1604) baery
Fuses mA/μA ranges: 0.8A/1000V ceramic fast blow A range; 10A/1000V
ceramic fast blow.