User Manual

24
13. Frequency
14 Peak Hold
Specied accuracy ± 150 digits.
Accuracy of Square Wave is unspecied.
15 VoltSeek
Voltage Range of High Sensivity : 80V ~ 1000V (At the top edge of the meter)
Voltage Range of Low Sensivity : 160V ~ 1000V (At the top edge of the meter)
Minimum Sensivity : > 5V (for ACV 1Hz ~ 10kHz)
> 20.0V (for ACV 10kHz ~ 50kHz)
unspecified (for ACV 50kHz ~ 100kHz)
Minimum Frequency : 1Hz
Overload Protection : AC/DC 1000V
Range OL Reading Resoluon Accuracy
100.00Hz 100.00Hz 0.01Hz ±(0.1% + 2D)
1000.0Hz 1000.0Hz 0.1Hz ±(0.1% + 2D)
10.000kHz 10.000kHz 0.001kHz ±(0.1% + 2D)
100.00kHz 100.00kHz 0.01kHz ±(0.1% + 2D)
16. Temperature
Does not include accuracy of the thermocouple probe. Accuracy specicaon assumes
surrounding temperature stable to ±1°C. For surrounding temperature changes of ± 2°C,
rated accuracy applies aer 2 hours. Overload Protecon: AC/DC 1000V
Range OL Reading Resoluon Accuracy
-40.0°C – 400.0°C 440.0°C 0.1°C ±(1% + 20D)
-40.0°F – 752.0°F 824.0°F 0.1°F ±(1% + 36D)