USER MANUAL – BEDIENUNGSANLEITUNG RND 320-KA3000 Series
English Safety Symbols This chapter contains important safety instructions that you must follow when operating the RND 320-KA3000 series and when keeping it in storage. Read the following before any operation to insure your safety and to keep the best condition for the RND 320-KA3000 series. Safety Symbols These safety symbols may appear in this manual or on the series..
RND 320-KA3000 — SERIES USER MANUAL Safety Instruction Safety Guidelines • Do not block or obstruct the cooling fan vent opening. • Avoid severe impacts or rough handling that leads to damage. • Do not discharge static electricity. • Do not disassemble unless you are qualified as service personnel. AC Input • AC Inut Voltage: 230 V, 50 Hz • Connect the protective grounding conductor of the AC power cord to an earth ground, to avoid electrical shock.
Fuse MODEL 220/230 V RND 320-KA3005D RND 320-KA3005P T3 A/250 V • To ensure fire protection, replace the fuse only with the specified type and rating. • Disconnect the power cord before fuse replacement. • Make sure the cause of fuse blowout is fixed before fuse replacement. Series Lineup MODEL RND 320-KA3005D RND 320-KA3005P V METER A METER USB RESOLUTION 4 DIGIT 4 DIGIT NO 10mV / 1mA Main Features Performance Low noise: cooling fan controlled by Heatsink temperature.
RND 320-KA3000 — SERIES USER MANUAL Front Panel Overview Over-Volt Protection Active Over-current Protection Active Constant Current Mode Constant Voltage Mode Output On/Off Indicator Setting Memory Indicator Display of voltage & current value Power Supply Indicator Front Panel lock_Out Function Over_current Protect On/Off Rapid Adjustment Knob for Voltage & Current Buttons for saving & Recalling Coarse and fine Voltage / Current Adjustment Buttons Buttons for saving & Recalling Voltage / Current Se
OCP is OCP indicator. When overcurrent function is turned on, indicator lights on. C.C is constant current indicator. When power supply is in the mode of constant current, this light is on. C.V is constant voltage indicator. When power supply is in the mode of constant voltage, this light is on. OUT is output indicator. If light on, there is voltage output in the output terminal.
RND 320-KA3000 — SERIES USER MANUAL Front panel LOCK function. For details, see Page 11. Over-Current protect on/off,. Pressing this button for more than 2 secends will make beep On/OFF. Over-votage Protect On/Off Output On/Off Voltage-Current Setting Adjustment Digit Selector Buttons Selection Voltage / Current for Adjustment Pressing the button, the volt indicator starts to blink; pressing it again, the ampere indicator starts to blink.
Fan Label on Voltage & Fuse USB Interface (only RND 320-KA3005P model) Power Socket RS232 Interface (only RND 320-KA3005P model) USB dependent interface bas ed on remote control order (see Page 14); only for 320-KA3005P model. RS232 dependent interface based on remote control order (see Page 14); only for 320-KA3005P model. The power cord socket mainly accepts AC values: 115V / 230V, 50 / 60 Hz. Please refer to the fuse parameters on the back fuse label to replace the specified fuse.
RND 320-KA3000 — SERIES USER MANUAL Operation Connect AC power cord Connecting AC power cord and selecting the corresponding AC voltage according to the back label on voltage; then connecting the AC power cord to the socket on the back panel. Power on3 Press the power switch to turn power on. The display initializes, showing the model of the machine and then showing the setting last. Power off Press the power switch again to turn power off.
Beep On / Off Panel Operation By default, the beep sound is enabled. To turn off he beep, press the OCP (BEEP) button for 2 seconds. A beep comes out and the beep setting will be turned off. To enable the beep, press the OCP (BEEP) button again for 2 seconds. Front Panel Lock Panel operation Press the LOCK button to lock the front panel button operation. To unlock, press the LOCK button for 2 seconds. Output Set Panel operation 1. Connecti ng the load to the front port, red( +),black ( -) . 2.
RND 320-KA3000 — SERIES USER MANUAL Contents The following list shows the setup contents: • Fine / coarse knob editing mode • Beep on / off • Output voltage / current level The following settings are always saved as «off»: • Output on / off • Front panel lock on / off Panel operation Press one of the 4 buttons (M 1,M2,M3,M4) and the LED light turns on accordingly. After you adjust the value, it is saved automatically once it stops blinking.
Remote Control for RND 320-KA3005P Remote Control Setup Only RND 320-KA3005P can be connected to the PC through interfaces USB/RS232 on the back of the machine and controlled by the remote control. COM setting Set up the COM port inside the PC according to the following list: • Baud rate: 9600 • Parity bit: None • Data bit: 8 • Stop bit: 1 Functionality check Run this query command via the terminal application such as MTTTY (Multi-threaded TTY).
RND 320-KA3000 — SERIES USER MANUAL Exiting from the Remote Control Mode 1. Close the remote control software. 2. Disconnect USB from the back. 3. The power supply disconnects; a tweet from the beep with the hint that the remote control is over. 4. The power supply automatically comes into the panel control mode. FAQ Q1: The panel buttons don’t work when power on. A1: The panel is locked. Press the button for over 2 seconds, and then the panel will unlock.
Specifications Note: The specifications below are tested under the conditions of temperation 25°C to -5 °C and the warm-up for 20 minutes. RND 320-KA3005P/D- Model VOLTAGE RANGE CURRENT RANGE 0 - 30 V 0- 5 A LOAD REGULATION VOLTAGE RANGE CURRENT RANGE ≤ 0.01% +2 mV ≤ 0.1% +10 mA LINE REGULATION VOLTAGE RANGE CURRENT RANGE ≤ 0.01% +3 mV ≤ 0.1% +3 mA SETUP RESOLUTION VOLTAGE RANGE CURRENT RANGE 10 mV 1 mA SETUP ACCURACY ( 25°C + -5°C) VOLTAGE RANGE CURRENT RANGE ≤ 0.5% +20 mV ≤ 0.
RND 320-KA3000 — SERIES USER MANUAL REACTION TIME VOLTAGE RANGE CURRENT RANGE ≤ 100 mS ≤ 100 mS INTERFACE RND 320-KA3005P RS232 USB ACCESSORIES USER MANUAL*1, POWER CORD*1 WEIGHT AND DIMENSION 110mm(W)* 156mm(H)* 260(D) The proprietary information in this manual is protected by copyrights. Any photocopies, reproductions or translation to another language are not allowed unless it is permitted. And all rights are reserved. The information in this manual is corre ct when printing.
Deutsch Sicherheitssymbole Diese Sicherheitssymbole können in diesem Handbuch oder in der Serie erscheinen.
RND 320-KA3000 — BEDIENUNGSANLEITUNG Inhalt 1. Beschreibung und Funktion........................................................................... 18 2. Bestimmungsgemäßer Einsatz...................................................................... 18 3. Sicherheits-, Service- und Betriebshinweise................................................. 18 4. Bedien-, Anschluss- und Anzeigeelemente................................................... 19 5. Inbetriebnahme und Betrieb.......................
1. Beschreibung und Funktion Das RND 320-KA3005D/P ist ein hochwertiges Labornetzgerät für den Konstantspannungsund Konstantstrombetrieb. Es verfügt über eine fein einstellbare Strom- und Spannungs einstellung, getrennte Anzeigen für Ausgangsstrom/ausgangsspannung, 5 Speicher für schnell abrufbare Strom-/Spannungseinstellungen sowie Schutzeinrichtungen gegen Überspannung/Überstrom.
RND 320-KA3000 — BEDIENUNGSANLEITUNG • • • • • • • • • Sicherheit zuständige Person beaufsichtigt oder erhielten von ihr Anweisungen, wie das Gerät zu benutzen ist. Das Gerät ist kein Spielzeug, es gehört nicht in Kinderhände. Es enthält heiße, kleine und verschluckbare Teile sowie Kabel, die Strangulierungsgefahren bergen. Lassen Sie Verpackungsmaterialien nicht achtlos liegen, diese können für spielende Kinder zu einer Gefahr werden.
Lufteinlass für Lüfter Sicherungstyp/ -größe Netzanschluss mit Sicherung USB-Port (nur RND 320-KA3005P Model) RS232-Anschluss (nur RND 320-KA3005P Model) Nur mitgeliefertes Netzkabel einsetzen! 5. Inbetriebnahme und Betrieb 5.1 Aufstellung • Stellen Sie das Gerät so auf, dass ringsum eine gute Luftzirkulation gewährleistet ist und es nicht längere Zeit eine Wärmequelle (Heizung, direkte Sonnenbestrahlung) ausgesetzt ist. 5.2 Inbetriebnahme • Trennen Sie eine eventuell angeschlossene Last vom Netzgerät.
RND 320-KA3000 — BEDIENUNGSANLEITUNG • Stellen Sie dann mit dem Adjust-Steller und den Fein-/Grob-Einstelltasten den jeweiligen Spannungs- oder Stromwert ein. • Das Gerät wechselt im betrieb automatisch zwischen Konstantspannungs- und Konstantstrombetrieb, jeweils angezeigt von den Anzeigen „CV/CC”. • Zunächst ist immer Konstantspannungsbetrieb (CV) eingestellt, bis der entnommene Strom die eingestellte Stromgrenze erreicht.
• Bei Überschreiten der Höchstspannung am Geräteausgang wird dieser nun sofort abgeschaltet. Die Anzeige „OVP” blinkt. • Nach Trennen der Last vom Gerät kontrollieren Sie dessen Funktion und stellen es durch erneutes Drücken der Taste „OVP” wieder zurück. Die Anzeige „OVP” verlischt. Überstromschutz • Schalten Sie den Ausgang ab und stellen Sie den gewünschten Höchststrom ein. • Drücken Sie die Taste „OCP”, die Anzeige „OCP” leuchtet auf.
RND 320-KA3000 — BEDIENUNGSANLEITUNG • Schalten Sie dann den Lastausgang mit dem Button „ON” (OUTPUT) rechts unten im Programmfenster ein. • Der Verlauf von Spannung und Strom wird automatisch in den beiden Grafikfenstern und in der Tabelle darunter aufgezeichnet. • Oben im Programmfenster kann der Überstromschutz des Gerätes nach Einstellung des Höchststroms zu- und abgeschaltet werden.