Datasheet
RND is a trademark of Distrelec
True RMS Multimeter
RND 355-00003
7. DC μA
8. AC μA
Input Impedance : Approx. 3kΩ
Overload Protecon : AC/DC 1000V
Range OL Reading  Resoluon  Accuracy
600.0μA  660.0μA  0.1μA  ±(1.0% + 2D)
Range  OL Reading  Resoluon  Accuracy
600.0μA  660.0μA  0.1μA  ±(1.5% + 3D)
LCD displays 0 counts when the reading < 10 counts.
Input Impedance : Approx. 3kΩ
Frequency Response : 45 ~ 500Hz (Sine Wave)
Ov
erload Protecon : AC/DC 1000 
9. Resistance
* There is a lile rolling less than ±50 digits when measuring >10.00 MΩ.
Overload Protecon : AC/DC 1000V
Range  OL Reading  Resoluon  Accuracy
600.0Ω  660.0Ω  0.1Ω  ±(0.9% + 5D)
6.000kΩ  6.600kΩ  0.001kΩ  ±(0.9% + 2D)
60.00kΩ  66.00kΩ  0.00kΩ  ±(0.9% + 2D)
600.0kΩ  660.0kΩ  0.1kΩ  ±(0.9% + 2D)
6.000MΩ  6.600MΩ  0.001MΩ  ±(0.9% + 2D)
40.00MΩ*  44.00MΩ  0.01MΩ  ±(1.5% + 5D)
Con
nuity : Built-in buzzer sounds when measured resistance is less than 20Ω and sounds 
o when measured resistance is more than 200Ω, Between 20Ω to 200Ω the buzzer maybe 
sound or o either.
Connuity Indicator : 2KHz Tone Buzzer
Response Time of Buzzer : < 500μsec.
Overload Protecon : AC/DC 1000V
11. Diode
10. Connuity
Range  OL Reading  Resoluon  Accuracy
600.0Ω  660.0Ω  0.1Ω  ±(0.9% + 5D)
Range  OL Reading  Resoluon  Accuracy
1.500V  1.550V  0.001V  ±(0.9% + 2D)
Open Circuit Voltage : Approx. 1.8V 
Overload Protection : AC/DC 1000V.