Datasheet
RND is a trademark of Distrelec
True RMS Multimeter
RND 355-00003
7. DC μA
8. AC μA
Input Impedance : Approx. 3kΩ
Overload Protecon : AC/DC 1000V
Range OL Reading Resoluon Accuracy
600.0μA 660.0μA 0.1μA ±(1.0% + 2D)
Range OL Reading Resoluon Accuracy
600.0μA 660.0μA 0.1μA ±(1.5% + 3D)
LCD displays 0 counts when the reading < 10 counts.
Input Impedance : Approx. 3kΩ
Frequency Response : 45 ~ 500Hz (Sine Wave)
Ov
erload Protecon : AC/DC 1000
9. Resistance
* There is a lile rolling less than ±50 digits when measuring >10.00 MΩ.
Overload Protecon : AC/DC 1000V
Range OL Reading Resoluon Accuracy
600.0Ω 660.0Ω 0.1Ω ±(0.9% + 5D)
6.000kΩ 6.600kΩ 0.001kΩ ±(0.9% + 2D)
60.00kΩ 66.00kΩ 0.00kΩ ±(0.9% + 2D)
600.0kΩ 660.0kΩ 0.1kΩ ±(0.9% + 2D)
6.000MΩ 6.600MΩ 0.001MΩ ±(0.9% + 2D)
40.00MΩ* 44.00MΩ 0.01MΩ ±(1.5% + 5D)
Con
nuity : Built-in buzzer sounds when measured resistance is less than 20Ω and sounds
o when measured resistance is more than 200Ω, Between 20Ω to 200Ω the buzzer maybe
sound or o either.
Connuity Indicator : 2KHz Tone Buzzer
Response Time of Buzzer : < 500μsec.
Overload Protecon : AC/DC 1000V
11. Diode
10. Connuity
Range OL Reading Resoluon Accuracy
600.0Ω 660.0Ω 0.1Ω ±(0.9% + 5D)
Range OL Reading Resoluon Accuracy
1.500V 1.550V 0.001V ±(0.9% + 2D)
Open Circuit Voltage : Approx. 1.8V
Overload Protection : AC/DC 1000V.