Datasheet
RND is a trademark of Distrelec
True RMS Multimeter
RND 355-00003
Electrical Specification
Accuracy is given as ±(% of reading + counts of least significant digit) at 23°C ± 5°C, 
with relative humidity Less than 80% R.H.
1. Tem
perature coecient
0.1 x (Specied accuracy) / °C, < 18°C, > 28°C
2. AC Funcon
ACV and ACA specicaons are ac coupled, true R.M.S. The crest factor may be up to
3.0 as 4000 counts. For square wave, Accuracy is unspecied.
For non-sinusoidal waveforms, Addional Accuracy by Cres Factor (C.F.):
Add 3.0% for C.F. 1.0 ~ 2.0.
Add 5.0% for C.F. 2.0 ~ 2.5.
Add 7.0% for C.F. 2.5 ~ 3.0.
3. DC 
mV
Input Impedance : 10MΩ
Overl
oad Protecon : AC/DC1000V
Range  OL Reading  Resoluon  Accuracy
600.0mV  660.0mV  0.1mV  ±(0.5% + 5D)
4. DC
 Voltage
Input Impedance : 10MΩ
Overload Protecon : AC/DC 1000V
5. AC Voltage
Range  OL Reading  Resoluon  Accuracy
6.000V  6.600V  0.001V  ±(0.5% + 2D)
60.00V  66.00V  0.01V  ±(0.5% + 2D)
600.0V  660.0V  0.1V  ±(0.5% + 2D)
1000V  1100V  1V  ±(0.5% + 2D)
Ran
ge  OL Reading  Resoluon  Accuracy
600.0mV  660.0mV  0.1mV  ±(1.0% + 5D)
6.000V  6.600V  0.001V  ±(1.0% + 3D)
60.00V  66.00V  0.01V  ±(1.0% + 3D)
600.0V  660.0V  0.1V  ±(1.0% + 3D)
1000V  1100V  1V  ±(1.0% + 3D)
LCD displays 0 counts when the reading < 10 counts. 
Input Impedance: 10MΩ // less than 100pF 
Frequency Response: 45~500Hz(Sine Wave)
Overload Protecon: AC/DC 1000V
6. Auto Voltage
LCD 
displays 0 counts when the reading < 10 counts.
Input Impedance: less than 3kΩ
Frequency Response: 45 ~ 500Hz (Sine Wave)
Overload Protecon : AC/DC 1000V
R
ange  OL Reading  Resoluon  Accuracy
600.0V  660.0V  0.1V  ±(2.0% + 3D)
1000V  1100V  1V  ±(2.0% + 3D)