Datasheet

AC+DC True RMS
Industrial Multimeter
RND 355-00002
RND is a trademark of Distrelec
(5)5) mA Current
Range
Resoluon Accuracy
60.00 mA
DC
0.01 mA ± (0.8% + 3d)
600.0 mA
DC
0.1 mA ± (0.8% + 3d)
60.00 mA
AC
0.01 mA ± (1.2% + 3d)
600.0 mA
AC
0.1 mA ± (1.2% + 3d)
Maximum measurement me: 10 minutes at 600 mA with at least 20 minutes rest me.
AC Frequency Response: 45 ~ 1 kHz (Sine Wave)
Overload Protecon: AC/DC 440 mA
(6)6) A Current
Range
Resoluon Accuracy
6.000 A
DC
0.001 A ± (0.8% + 3d)
10.00 A
DC
0.01 A ± (0.8% + 3d)
6.000 A
AC
0.001 A ± (1.2% + 3d)
10.00 A
AC
0.01 A ± (1.2% + 3d)
Maximum measurement Current: 20 A
Maximum measurement me:
>5 A for max.3 minutes with at least 20 minutes rest me.
>10 A for max.30 seconds with at least 10 minutes rest me.
AC Frequency Response: 45 ~ 1 kHz (Sine Wave)
Overload Protecon: AC/DC 11 A
(7)7 Peak Hold:
Specied accuracy ± 150 digits for < 6000 counts.
Specied accuracy ± 250 digits for >= 6000 counts
(8)Resistance
Range Resoluon Accuracy
600.0 Ω 0.1 Ω ± (0.8% + 5d)
6.000 KΩ 0.001 KΩ ± (0.8% + 2d)
60.00 KΩ 0.01 KΩ ± (0.8% + 2d)
600.0 KΩ 0.1 KΩ ± (0.8% + 2d)
6.000 MΩ 0.001 MΩ ± (0.8% + 2d)
40.00 MΩ* 0.01 MΩ ± (1.5% + 5d)
*There is a lile rolling less than ±50 digits when measuring > 10.00 MΩ.
Overload Protecon: AC/DC 1000 V
(9)9) Continuity
Range
Resoluon Accuracy
600.0Ω 0.1Ω ± (0.8% + 5d)
Connuity: Built-in buzzer sounds when measured resistance is less than 20 Ω and sounds
o when measured resistance is more than 200 Ω, Between 20 Ω to 200 Ω the buzzer
maybe sound or o either.
Connuity Indicator: 2 KHz Tone Buzzer
Response Time of Buzzer: < 500 μsec.
Overload Protecon: AC/DC 1000 V