REV. 1.10/2002
RMT LtdRMT Ltd
RMT LtdRMT Ltd
DX6106
2-6
Calculated coefficients … of polynomial
expression and zero ratio may be stored into device
internal on-board EEPROM memory.
The first calibration is made by Manufacturer.
The factory standard calibration uses not less than 5
standard gas mixtures.
Several calibrations as above described are made at
different ambient temperatures (in a specified
operating range) and at corresponding optimal
operating temperatures of integrated detector-emitter
pair.
Up to 15 such calibrations are possible to store for
further application.
Format of calibration data stored in the EEPROM
memory chip after manufacturer calibration is
described below in Chapter 3.
2. During routine operation the detector’s output
signals should be measured to calculate ratio. Using
known “zero” ratio value is calculated.
Finally, using known polynomial coefficients …
gas concentration may be calculated with high
accuracy.