Calibrating Instruments with Standard Sensors • 36
To perform a calibration on an instrument:
1. Confirm that the AC Adapter is connected to the SDM-E2 and to an AC wall
socket.
2. Press and hold the SDM-E2’s POWER button. The LEDs will turn amber.
3. When the BUMP T and CAL S LEDs turn off, release the POWER button.
4. The COPY LED will be off or on steadily and the CHARGE LED will be blinking
green if the SDM-E2 is operating properly or solid red if there is a system
failure. The amount of free memory in the SDM-E2 will dictate the condition of
the COPY LED (see “Available Memory in the SDM-E2” on page 68).
5. Install an EAGLE 2 in the instrument cradle. Set the EAGLE 2 onto the top of
the SDM-E2 and slide the instrument forward until it falls into place as
illustrated below.
Figure 12: Inserting the EAGLE 2
WARNING: Inserting the EAGLE 2 improperly may damage the exhaust bellow at the
back of the SDM-E2.