16 • Setting the Operational Parameters in Edit Mode
Turning on the SDM-2009 with an Instrument
Do the following to turn on the SDM-2009 and establish a connection with an
instrument:
1. Confirm that the AC adapter is connected to the SDM-2009 and to an AC wall
socket.
2. Press and hold the SDM-2009’s POWER button. The LEDs will turn amber.
3. When the BUMP T and CAL S LEDs turn off, release the POWER button.
4. The COPY LED will be steadily on or off and the CHARGE LED will be blinking
green if the SDM-2009 is operating properly or solid red if there is a system
failure. The amount of free memory in the SDM-2009 will dictate the condition
of the COPY LED (see “Available Memory in the SDM-2009” on page 38).
5. Install an instrument into the instrument cradle.
6. Press and hold the POWER MODE button on the instrument until you hear a
beep, then release it. The instrument will begin its power up sequence (see
exceptions below in step 7). If a successful connection between the instrument
and the SDM-2009 occurs, the home screen will appear on the instrument
display at the end of the startup sequence. The auto calibration values for all
installed sensors will be displayed.
NOTE: The screen shown above applies to a 4-channel GX-2009. If your
GX-2009 has less than 4 channels, the inactive channels will not
appear in the above screen or in any screens where channels are
displayed. All screens in this manual assume a 4-channel GX-2009.
7. There are two exceptions to the sequence described in step 6 above. See the
GX-2009 User Setup Program Operator’s Manual for a description of the Cal
Limit Display and Cal Limit Check instrument parameters.
When Cal Limit Display is set to On and Cal. Limit Check is set to Can’t
Use, if the GX-2009 is due for calibration the instrument will show a
calibration expired failure screen and the buzzer and LED arrays will pulse
for a few seconds.
The instrument will then connect to the SDM-2009 and display the home
50 12.0
TRANSMIT
CO ppm H2S ppm
CH4 %LEL OXY vol%
50 25.0
FAIL
C-LIMIT