User Manual

PT2E-1140
Display Part
MAINTENANCE/ESC switch‥‥Used for entering into the maintenance mode with the
control key. And used for cancel the maintenance mode.
PW/TR light‥‥……………… Illuminates continuously when the equipment is working
(power light). And flickers in the case of abnormality
in the equipment.
AL1 light ……………………‥‥ Illuminates when 1
st
alarm is activating.
AL2 light …………………‥‥ Illuminates when 2
nd
alarm is activating.
⑤ SKIP light…………………‥‥ Illuminated when point skip is selected. And flickers in the
maintenance mode.
AL-T(▲) switch …………‥‥ Used to increase the value with the control key.
LCD ‥‥………………………. Indicates the gas concentration and error code.
AL/SET switch ……………… Used for confirmation of preset alarm level. And used for
decision in the maintenance mode.
⑨ ZERO light …‥‥………………Flickers in the zero adjustment mode.(Steady light when
the adjustment is over.)
SPAN light …‥‥……………… Flickers in the span adjustment mode. .(Steady light when
the adjustment is over.)
mA light……………………‥ Illuminates with the current output indicated on the LCD
(during maintenance)
mA (▼)switch‥‥……………… Used for indicating current output on the LCD. And used to
decrease the indication in the maintenance mode.
4