User Manual

PT2E-1140
LCD indication Menu
Displays set potential
Zero adjustment mode
Span adjustment mode
4mA adjustment mode
Alarm point set mode
Alarm test mode
Point skip set mode
<Common operation ∼ To enter maintenance mode>
To enter maintenance mode, press MAINTENANCE switch● by control key for
3 seconds in normal measuring mode.
When entered maintenance mode after 3 seconds, 4-20mA output becomes 2.5mA
and SKIP light is flickering. Release control key after entered.
Change indication number with pressing UP▲ or DOWN▼ switch by control key.
And press SET switch● when the mode you want to enter is indicated.
<Common operation∼To change the mode>
Display goes to MENU when press MAINTENANCE switch● during each mode.
Change indication number with pressing UP▲ or DOWN▼ switch by control key.
And press SET switch● when the mode you want to enter is indicated.
<Common operation∼To cancel maintenance mode>
Display goes to MENU when press MAINTENANCE switch● during each mode.
To recover measuring mode, press MAINTENANCE switch● for 3 seconds during
MENU
When entered measuring mode, SKIP light goes off and indicating as it is measuring mode.
4−3 Gas Sensitivity Calibration Method
The following tools and jigs are necessary for zero and span adjustments.
Gas sampling bag
Calibration adaptor
Calibration gas
Control key
20