User Manual

SC-01 Operator’s Manual Specifications • 2
Specifications
Table 1: 01 Series Specifications
Available Target Gas/
Detection Range/
Display Increment
Ammonia (NH
3
): 0 - 75.0 ppm, 0.5 ppm increments
Arsine (AsH
3
): 0 - 1.50 ppm, 0.01 increments
Carbon Monoxide (CO): 0 - 75.0 ppm, 0.5 ppm increments
Chlorine (Cl
2
): 0 - 3.00 ppm, 0.02 ppm increments
Hydrogen Cyanide (HCN): 0 - 15.0 ppm, 0.1 increments
Hydrogen Sulfide (H
2
S): 0 - 30.0 ppm, 0.2 ppm increments
Phosphine (PH
3
): 0 - 1.00 ppm, 0.01 ppm increments
Sulphur Dioxide (SO
2
): 0 - 6.00 ppm, 0.05 ppm increments
Detection Principle Electrochemical
Sampling Method Diffusion
Response Time 90% in 60 seconds
Accuracy ±10% of reading or ± 5% of detection range full scale
(whichever is greater)
Indication Digital LCD
Safety/Regulatory
C US
186718
CSA classified, “C/US”, as Intrinsically Safe. Exia. Class I,
Groups A, B, C, & D. Temperature Code T3.
Power Source Two AA size Alkaline Batteries Standard, Duracell MN1500 or
PC1500
Continuous
Operating Hours
At 25 °C: Over 250 Hours, No Alarms or Backlighing
Case High-impact Plastic, Dust and Weather Proof
Standard
Accessories
Alligator Clip
Rubber protective boot
Optional Accessories Calibration Cup
Calibration Kit
Belt Clip
10 ft. Extender Cable
Data Logging Kit
Dimensions and
Weight
2.5” W x 5.2” H x 1.2” D (63 mm x 131 mm x 31 mm)
7.6 ounces (215 g)