User Manual

23 • Setup Mode SC-01 Operators Manual
1. When the DATE Screen is displayed, press and release the POWER
button. A screen appears with the year flashing in the upper left, the
date in the middle, and the time in the lower right.
2. Use the AIR or the DISP button to display the desired year.
3. Press and release the POWER button to save the setting. The month
setting flashes.
4. Repeat steps 2 and 3 to enter the month, day, hour and minute setting.
When you save the minute setting, you return to the DATE Screen.
Performing a Fresh Air Adjustment
Entering the AIR CAL menu item allows you to perform a fresh air
adjustment.
Perform a fresh air adjustment when you are performing a calibration
before proceeding to the next menu item, SPAN CAL, to perform a span
adjustment.
WARNING: Calibrate the SC-01 in a non-hazardous environment.
1. Find a fresh air environment, an environment of normal oxygen content
(20.9%) that is free of toxic or combustible gases.
2. When the AIR CAL screen is displayed, press and release the POWER
button. A screen appears that displays the current gas reading and
prompts you to press the AIR button with a flashing “PUSH AIR”
message at the top of the screen.
2006
12.11
08:00
AIR CAL
PUSH AIR
0.00
PPM