PT0E-0730 RIKEN PORTABLE COMBINATION GAS DETECTOR MODEL RX−516 RX−517 OPERATION MANUAL MARINE USE Precaution in operation ・After reading this manual well, start to operate this instrument. ・Keep this manual where to carry out easily ・When take out this manual by use of construction work, be sure to put it where it was. ・Do not use this instrument for no other than given purpose.
PT0E-0730 In the beginning It is of our great pleasure to purchase Riken portable combination gas detector model RX-516/RX-517 this time. This instrument is an explosion-proof type portable gas detector which is designed to measure the presence of the following gas at crude oil tankers shown below. RX-516 • Hydro carbon(HC) • Oxygen(O2) • Hydrogen sulfide(H2S) :0-100.0%LEL/2-100.0vol% (Iso-butane cal.) :0-25.0 vol% :0-100.0ppm RX-517 • Hydro carbon(HC) • Oxygen(O2) • Hydrogen sulfide(H2S) :0-100.
PT0E-0730 C O N T E N T S 1.Caution in operation ・・・・・・・・・・・・・・・・・・・・・・・・・・・・・ 3 2.Name of each part・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・ 4 3.Operation・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・ 8 3−1.Check item before operation・・・・・・・・・・・・・・・・・・・・・ 8 3−2.Preparation・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・ 8 3−3.Start-up and end-up・・・・・・・・・・・・・・・・・・・・・・・・・・ 9 3−4.Flow monitor check・・・・・・・・・・・・・・・・・・・・・・・・・・ 10 3−5.Zero point adjustment・・・・・・・・・・・・・・・・・・・・・・・・ 10 3−6.Operational procedure・・・・・・・・・・・・・・・・・・・・・・・・ 12 3−7.Peak value
PT0E-0730 1.Caution in operation To maintain the function for safety, follow the following instruction. Warning ・Do not modify nor change the circuit and structure etc. When modify or change, it could not maintain the function. ・This is explosion-proof product. Battery has to be replaced in non-hazardous zone. The battery replacement in hazardous zone shall be out of the scope for explosion-proof product. ・This is explosion proof. Be sure to operate with carrying case on.
PT0E-0730 2.Name of each part ① ⑥ ② ③ ④ ⑤ GAS OUT H2S Low・HC・O2 GAS IN ⑦ ⑧ GAS OUT H2S high ①POWER/ENTER SW This is used for power ON/OFF and entering the figure. ②PUMP/− SW This is used when make the pump ON/OFF and deverse the figure. ③ZERO/+SW This is used when make zero adjustment and increase the figure. ④PEAK SW This is used when to let the peak displayed. ⑤MODE/ESC SW This is used when enter the set mode or cancel the input.
PT0E-0730 《Display of RX-516/517》 Hydro carbon「HC」 Oxygen「OXY」 RX-517 Hydrogen sulfide(High) 「H2S」 RX-516/RX-517 Hydrogen sulfide(Low) 「H2S」 *Caution ・For HC, when it goes over 100%LEL, it shall be changed to vol% automatically. 《RX-517 case》 ・ H2S can be changed over by H2S gas range mode cock.
PT0E-0730 【Standard accessories】 ①Carrying case with shoulder strap ②Filter tube with flow monitor and junction tube ③Gas sampling probe ④Gas sampling hose ⑤Alkaline battery 6
PT0E-0730 【Optional accessories】 ①Gas sampling hose 30m with plummet ②Gas sampling hose 30m with float type sampling head ③Gas sampling bag 1L for LEL and VOL and calibration gas cylinder ⑤Absorbent cotton ⑥Water proof filter for filter tube (10pcs.
PT0E-0730 3.Operation 3−1.Check item before operation RX-516 or 517 instrument: ・Check that there is no damage for display unit etc. Filter tube with flow monitor: ・Check the dirt of inside cotton filter. When it gets dirt, replace the cotton filter with new one. ・When water drops remain, wipe it out and change it with new one. ・Check that these is no crack or damage. ・Check that these is no loose mount on the nipple. Gas sampling hose: ・Check that there is no crack, creek or twist on the instrument.
PT0E-0730 3−3.Start-up and end-up When press power switch for 3 sec, the power gets on and this shall start up sequencing the program of the voltage check, filter connection check and sensor connection check respectively. Press power switch for 3 sec.
PT0E-0730 3−4.Flow monitor check When the power gets ON, the pump shall start operation check that suction works properly by filter tube with flow monitor. Float At normal sucking(Can see float) Not sucking(Can not see float) 3−5.Zero point adjustment Zero point calibration for this instrument shall be done HC, O2(Oxygen), H2S(Hydrogen sulfide). Caution When make zero adjustment, proceed to do it after having fresh air sucked into instrument for over 1 min.
PT0E-0730 <FLOW CHART OF ZERO POINT ADJ> ZERO MODE DISPLAY ZERO key ・Press ZERO key for 3 sec. HOLD KEY ・When press ENTER key, HC, O2, H2S zero adj ESC key Move to zero adj shall start. FRESH AIR? ENTER key for CO2 adj ESC key ADJ ZERO OK ・Calibration 《Zero set OK》 《Zero adj fails》 FAIL ZERO ZERO SET OK Press for 5 sec and move to the display. All are OK HC fails *Note ・To cancel previous operation, press ESC key. But can not cancel after press ENTER key.
PT0E-0730 3−6.Operational procedure Approach the sampling hose to the sampling site for measurement and leave sampling gas sucked. ・Take care that any twist on the sampling hose can not be made. ・Stop sampling hose at the point of measurement. ・From the hose length consider the delay time and read out the reading after the reading gets stable. Danger ・Oxygen depletion air etc may be exhausted from gas outlet. Do not breathe in there absolutely. ・High density gas over (LEL) may be exhausted.
PT0E-0730 3−7.Peak SW value mode When press PEAK SW during operation, the peak valve from power on to the current point (HC, H2S = MAX, O2=MIN) shall be shown. 【RX-516/517】 Peak mode Operation mode PEAK SW P E A K 【RX-517】 H2S change over H2S high display Peak display PEAK key P E A K *Note 《RX-517 case》 ・In the H2S high display, the peak value of H2S is displayed. ・When changed from peak display, it changes to the operation display.
PT0E-0730 Operation mode PUMP key ・Press pump key for 3 sec. HOLD KEY ・Pump work will be off. PUMP OFF PUMP key ・Press pump key for 3 sec. HOLD KEY PUMP ON ・Pump work will be on. Hold and press for 5 sec and back to the operation display. 3−9.Operation finished (1)Treatment after operation Wind up sampling hose so treat it can not be twisted and bent. Supply fresh air in the condition of sampling hose connected and make air-cleaning until the reading gets to zero nearly.
PT0E-0730 3−10.Practical application (1) Tank inside of crude oil tanker By use of sampling hose with plummet, it is available to monitor the gas density in the tank. 《Sample in crude oil tanker》 COMBINATION GAS DETECTOR AIR VENT HEAD GAS SAMPRING PIPE GAS SAMPRING PIPE WBT Caution When measure O2 in insert gas for a long time, be sure to use CO2 in the atmosphere of below 15%. When use higher density than 15%, make the measuring time as short as possible.
PT0E-0730 《Interference measures in inert gas》 In the component of inert gas, a lot of interference gas may be present. When it should be present so much to affect the measurement, it shall be required to mount the interference removal filter. 4.Alarm 4−1.Alarm display ①At 1st alarm, the figure will blink and “Warring” will blink. ②At 2nd alarm, the figure will blink and “Alarm” will blink. ③At over alarm, the figure will blink and “Over” will blink.
PT0E-0730 4−3.Trouble display (1)Low flow alarm ・When low flow trouble triggers, “LOW FLOW” is displayed and pump stops working. ・Buzzer sounds continuously. ・After checking the cause, reset low flow alarm by “ESC” key. The pump will re-start and return to the operation display. Pump ON Pump OFF LOW FLOW ESC key Caution When low float trouble took place, be sure to check the cause to make low flow. Especially when suck water, it will be the cause for trouble.
PT0E-0730 (3)System trouble alarm ・When the program is troubled, “SYSTEM FAIL” will be shown. ・Buzzer will sound continuously. ・After “SYSTEM FAIL” is shown, it does not accept other than “Power off”. SYSTEM FAIL S A V E Power off available only by power SW (4)Sensor trouble alarm ・When sensor is troubled, “FAIL SENSOR” is displayed and the measuring gas for the troubled will be shown. (Sensor for trouble detected at start-up time shall be shown.) ・Buzzer will sound continuously.
PT0E-0730 5.Each function 5−1.Operation mode control Operation mode ・Press MODE key for 3 sec. MODE key HOLD KEY Displayed in the following order ① 1st alarm ①1st alarm display ② 2nd alarm MODE key ③ Data display ④ Data logger remaining W A R N I N G − + PEAK key To alarm setting − ・ 『+』or『−』 key: When press it, the display will be changed.
PT0E-0730 5−1−1.Alarm setting 1st alarm display ・Input the password【3002】. ・When press『+』or『−』, the number will be changed. W A R N I N G ・Fix the number by ENTER key and move to next PEAK key digit. ・After fixing the last digit, move to alarm setting (When make the password error, it will show “ERROR”. Press ESC key.) M ‐ P A S S W O R D ENTER key ・First, 1st alarm point shall be shown on the display. ・Change 『+』or『−』key and confirm it by pressing ENTER key.
PT0E-0730 5−1−2.Date/time setting Date/time display D A T E / T I M E ・First, set the year. ・The set number shall blink. PEAK key ・The change of number is set by 『+』 or『−』key. ・ Press ENTER key and the number is ESC key D A T E / T I M E entered. ・When press “ESC” key, the set work stops in the way and go back to the date ENTER key /time display. Year Month Day Set each in this turn. Hour Minute ・After finish the minute, press “ENTER” key.
PT0E-0730 5−3.Data logger Data logger function has following 5 kinds. (1)Interval trend The change of measuring gas shall be recorded until the power on into off. Recording time can 100 times data. When it is over 100 times, the old data shall be deleted and record the new data by up-dating. ※But when exceed the max recording time within 100 times, the old data shall be deleted. The max record time for interval time shall be as follows.
PT0E-0730 6.Maintenance check For this correct operation, follow to the following procedure. As this is safety instrument. Be sure to make the regular maintenance. If any trouble should take place, contact RIKEN KEIKI or nearest agent. 6−1.Battery replacement Warning ・ For battery replacement, make it non-hazardous area by all means after turning off the power. ・ Be sure to change the battery with the genuine ones. When replace the batteries, replace all four pcs with new ones.
PT0E-0730 6−2.Gas calibration (Span adjustment) ・ Hold and pressing ZERO and PUMP key similtaneously in the operation mode. ZERO key PUMP key HOLD KEY ESC key + SPAN GAS HC ENTER key ・ First, CH4 display shall be shown. ・ When press EMTER key, “HC” gas calibration shall start. ・ When press 『+』key, it shall change to gas calibration of 「OXY」→「H2S」. ・ When pressing ESC key, it shall return to “Operation display”. ・This shall display gas name, gas density and unit.
PT0E-0730 Danger ・Compressed cylinder gas is used for gas calibration and take care at handling. ・At calibration by combustible gas, do not approach it to the fire. ・Exhaust the remaining gas in the sampling gas appropriately. Caution Before gas calibration, be sure to make zero adjustment. 《Handling of sampling bag and canned gas》 ① Sample the cal gas into sampling bag from canned gas. Canned gas Sampling bag ② Sampling bag with cal gas in shall be used by putting to gas in side at gas calibration.
PT0E-0730 《Cal gas make-up by H2S gas checker CK-82》 ① ② ③ ④ Take out 1 pce. ampoule from accessory ampoule pack. Check that the rubber caps are fixed in the cover of container. Open the cover of container and deflate the air by squeezing the pressure adj bag. Put up the ampoule crusher know and put the ampoule into ampoule crusher inside of the cover shown below. When the ampoule is put in reversely, the ampoule may be crushed.
PT0E-0730 6−3.Filter replacement For filter (cotton) inside of filter with flow monitor, check it before use and when it gets dirty, replace it. (1)Turn the knurling part in depth for filter tube with flow monitor and open the inside. (2)Remove the cotton of it by tweezers etc. (3)Put the cotton filter uniformly on even base inside of filter tube and return back the filter tube in reverse procedure. Knurling cap Caution ・ Do not put cotton filter too much inside.
PT0E-0730 6−6.The replacement of parts The following parts have a life and each, regular replacement shall be recommended. When replace, contact RIKEN KEIKI or nearest agent. 《Recommended consumable parts list for replacement》 No.
PT0E-0730 6−7.Storage or treatment not to use for a long period The instrument shall be kept in the following condition. Temp :10∼30℃ Humidity :30∼80%RH Environment:Place free from gas and solvent vapours. Store the instrument in the box kept in originally. When the storage box was lost, store it in vinyl pack etc.
PT0E-0730 8.Trouble shooting This trouble shooting does not mention all the trouble causes. The following is mentioned briefly by picking up the point to help survey of trouble cause in high frequency. Phenomenon SYSTEM FAIL Cause ・Excess noise effects etc. FAIL SENSOR ・Sensor trouble ・Did excess shock etc by dropping, crushing. ・Water, oil are soaked. FAIL FLOW LOW BATTERY (REPLACE BATT.
PT0E-0730 9.Definition of words Vol% Gas density is expressed by the unit of a part. ppm Gas density is expressed by the unit of a part per. 10,000ppm=1vol% %LEL Suppose to put the Lower Explosive Limit as 100 and at that time, combustible gas density is expressed by the unit of a part per hundred. Iso-butane:100%LEL=1.8vol% N-hexane :100%LEL=1.2vol% Calibration By use of calibration gas etc, it is to adjust the reading, display value or set value to the true valued.
PT0E-0730 10.Specifications Refer to the specifications for separate catalog overleaf. Model RX-516 (HC/O2/H2S 0-100.0ppm) , RX-517 (HC/O2/H2S 0-100.0ppm & 0-1000ppm) Gas detected Iso-butane (HC) Oxygen (O2) Hydrogen sulfide (H2S) Detection principle NDIR(Non-Dispersive infrared) Galvanic cell Electro-chemical cell Detection range (digit) 0-100%LEL (0.5%LEL) 2-100vol% (0.5vol%) 0-25.0vol% (0.1vol%) Operating Temp. & Humidity 0-100.0ppm (0.
PT0E-0730 11.Detection principle 11−1.NDIR method (Non-Dispersive Infrared) The model is based on NDIR method (Non-Dispersive Infrared) and this structure is shown below. The infrared beam emitted from the light source passes through the measuring cell, and optical band pass filter which can pass the absorption wave of measuring gas and attains to the infrared sensor. The amount of infrared attaining to the infrared sensor through the measuring cell and will decrease according to its density.
PT0E-0730 11−3.ELECTRO-CHEMICAL CELL method This gas detector applies an electrochemical sensor. The sample gas is electrolyzed by the electrolyzed cell added with specific electric potential (bias voltage) and detected from the electrolyzed current generated at that time. The electrochemical sensor is designed to keep the interface between electrode and electrolyte at a constant potential (Bias voltage) and is the method to electrolyze gas directly.