Instruction Manual

7 Maintenance 7-2. Regular maintenance mode
RM-5000 - 64 -
<Environmental Setting 3 "2-7">
In the environmental setting 2, specify the settings of functions. (* It is recommended that setting changes
should be recorded in a log.)
The environmental setting 2 includes setting menus which are usually not used. Be careful not to change
these settings by mistake.
<<Environmental Setting 2>>
2-7. SETTING3
Press SET key.
(6 seconds later)
SE 0. AMP DEF
Initialize the amplifier PCB.
To initialize the amplifier PCB, press
the SET key.
(Press the MODE key to cancel this
menu.)
SE 1. HEAT ADJ
A
djust the heater
current.(GP-5001, NC-5001,
NC-5001W, GH-5001 only)
Press SET key to confirm the
selection.
SE 2. LOAD ADJ
A
djust the load
voltage.(GH-5001 only)
Press the or key to
change the value, and then
press the SET key to confirm
the value.
During maintenance
During inhibit
Lamp on
Lamp off
Lamp blinking
(3 seconds later)
(3 seconds later)