Instruction Manual
7 Maintenance 7-2. Regular maintenance mode
- 61 - RM-5000
SE 12. MNT OUT
Set the external output for the
maintenance mode.
Press the ▲ or ▼ key to
select either 2.5 mA/4.0 mA
/HOLD (previous value)/4 - 20
mA (linked to display value),
and then press the SET key to
confirm the selection.
SET 13. MA 4-20
A
djust the external output (4 -
20 mA).
A
fter selecting 4 mA or 20 mA,
press the SET key to enable
adjustment.
A
fter adjustment of 4 mA is
completed, perform the
adjustment of 20 mA.
Press the ▲ or ▼ key to
select the output (%), and
then press the SET key to
confirm the value.(It must be
adjusted to the upper unit or
the ammeter. Press the
MODE key to skip this menu
during the operation.)
To SE 0. ADDRESS