Combustilbe Gas Transmitter 26 Pioneer Operators Manual (PRELIMINARY)
8. Verify that the Zero Proceed line is ashing [or Zero (Oxygen Span) if applicable],
then press the ENTER button to continue. If the Zero Proceed line is not ashing,
press the UP/YES button, then press the ENTER button.
The Fresh Air screen appears. The readings displayed on the Fresh Air screen are the
current gas readings. The asterisk (*) to the right of each reading indicates that you
cannot adjust this reading at the Pioneer.
CAUTION: Do not press the ENTER button at this time.
Setting the Zero Reading
1. Screw the regulator into the zero air calibration cylinder.
2. When the reading on the voltmeter stabilizes, verify a reading of 100 mV (±2 mV).
If necessary, use the zero potentiometer on the amplier to adjust the reading to
100 mV (±2 mV).
3. Unscrew the regulator from the zero air calibration cylinder.
Leave the sample tubing connected to the regulator and the calibration cup.
Setting the Response Reading
1. Screw the regulator into the calibration cylinder.
2. When the reading on the voltmeter stabilizes, verify that the reading matches the
response reading (±2 mV) you determined earlier. If necessary, use the span
potentiometer on the amplier to adjust the reading to match the correct response
reading.
3. Unscrew the regulator from the calibration cylinder.
4. Remove the voltmeter leads from the amplier test points.