User Manual

Pioneer Operators Manual (PRELIMINARY) Combustible Gas Transmitter 21
Slow or No Response/Difficult or Unable to Calibrate
Symptoms
The transmitter responds slowly or does not respond during the monthly response
test.
Unable to accurately set the zero or response reading during the calibration
procedure.
The transmitter requires frequent calibration.
NOTE: Under normal circumstances, the transmitter requires calibration once every
three months. Some applications may require a more frequent calibration
schedule.
Probable causes
The calibration cylinder is low, out-dated, or defective.
The transmitter is malfunctioning.
Recommended action
1. Verify that the calibration cylinder contains an adequate supply of a fresh test sample.
2. If the calibration/response difculties continue, replace the detector as described later
in this section.
3. If the calibration/response difculties continue, contact RKI Instruments, Inc., for
further instruction.
No Display Reading
Symptoms
The transmitter is installed, but the display screen is blank for that channel.
NOTE: If the display screen is blank for all channels, see the troubleshooting guide in
Chapter 5, Maintenance.
Probable causes
The channel status setting is set at NO SENSOR/AMP.
Recommended action
1. Enter the Channel Setup program, and change the channel status setting to ACTIVE/
ONLINE or SILENCED/OFFLINE.
2. If the display reading difculties continue, contact RKI Instruments, Inc., for further
instruction.