User Manual

24 • Operation Pioneer Gas Monitor Operator’s Manual
Running the Instrument Setup Program
This section describes how to enter the program, display and change program settings,
and save or cancel changes to the program settings.
Entering the instrument setup program
1. Open the housing door, and locate the program buttons to the right of the display
screen.
2. Press the ENTER and ESCAPE buttons simultaneously three times.
The display screen indicates that you are in the Instrument Setup program.
3. Press the ENTER, UP/YES, or DOWN/NO button to continue.
The program displays the Alarm ON Delay screen.
Alarm on delay screen
This screen indicates the length of time that the Pioneer delays alarm indications (buzzer
and relays) when it senses an alarm 1, alarm 2, or alarm 3 condition. You can choose from
a range of 0 Seconds to 60 Minutes. The default setting is
0 Minutes, 1 Seconds
.
To accept the alarm on delay setting and continue the program, press the ENTER
button.
The program displays the Alarm OFF Delay screen.
To update the alarm on delay setting:
1. Use the UP/YES and DOWN/NO buttons to display the setting you want.
2. Press the ENTER button to continue the program.
The program displays the Alarm OFF Delay screen.
Alarm off delay screen
This screen indicates the length of time that the Pioneer delays turning off alarm
indications (buzzer and relays) after an alarm 1, alarm 2, or alarm 3 condition passes. You
can choose from a range of 0 Seconds to 60 Minutes. The default setting is
0 Minutes,
0 Seconds
.
To accept the alarm off delay setting and continue the program, press the ENTER
button.
The program displays the Relay Action screen.
To update the alarm off delay setting:
1. Use the UP/YES and DOWN/NO buttons to display the setting you want.
2. Press the ENTER button to continue the program.
The program displays the Relay Action screen.
Alarm
O
ND
e
l
ay:
0
Minutes
1
S
econds
Alarm
O
F
De lay :
0
Minutes
0S
econds
F