Manual

16 Alarms Eagle Instruction Manual
STEL Alarm (Toxics Only)
If a toxic gas channel’s average gas reading for the past 15 minutes
exceeds the STEL alarm setting:
STEL displays in the alarm field for that channel.
The channel’s display line flashes.
The buzzer sounds a pulsed tone.
The alarm lights flash.
TWA Alarm (Toxics Only)
If a toxic gas channel’s average gas reading for the past 8 hours
exceeds the TWA alarm setting:
TWA displays in the alarm field for that channel.
The channel’s display line flashes.
The buzzer sounds a pulsed tone.
The alarm lights flash.
Over Range Alarm
If a channel’s gas reading exceeds that channel’s full-scale setting:
OVER displays in the alarm field for that channel.
The channel’s display line flashes.
The buzzer sounds a pulsed tone.
The alarm lights flash.