User Manual
Viewing, Printing, Exporting, & Deleting Data in the Data Window • 29
the program window and select the data you want to view.
Calibration History
The instrument is capable of saving information for up to 124 of the most 
recent bump tests and calibrations combined. This calibration history is 
retrieved by the Data Logger Management Program when data is 
downloaded from the instrument using either the Complete Download 
download command or the Download Logger Data download command.
Instrument calibration information is also available in the Last Calibration 
Window. The information regarding the most recent successful calibration 
and bump test for each downloaded instrument along with whether or not 
an instrument is due for calibration or bump testing can be accessed using 
the Last Calibration button on the right side of the program window. See 
“Last Calibration Window” on page 64 for a complete description of the 
information that can be accessed by the Last Calibration button.
All downloaded calibration information for all downloaded instruments is 
available in the Calibration History folders in the Data Window. This 
information is saved in a Calibration History Folder that is located in an 
untitled folder for each instrument. The calibration history files are 
differentiated by instrument. The calibration information available here is 
more comprehensive than that in the Last Calibration Window. The 
calibration information for all calibrations downloaded, whether successful 
or not, is saved instead of just the most recent successful calibration for 
each instrument.