User Manual
10 • Control Buttons
Data Button
Clicking the Data button opens the Data Window.
In the Data Window, you can view, print, export, and delete data that has 
been downloaded from instruments. 
NOTE: If the GX-2012 Data Logger Management Program database is 
on the user’s computer (default setting), data will only appear for 
instruments downloaded to that computer. If the database is on a 
network, data will appear for all instruments downloaded to that 
database. See “Accessing a Common Database” on page 14 for 
more instruction.
The following types of data files are saved in the Data window:
• Bump Test Files
A bump test file is saved for each instrument that has been 
downloaded. It records the bump test information for every bump test 
that was downloaded. The instrument can save information for up to 
124 bump tests and calibrations combined in its memory.
 Figure 6: Data Window