Instruction Manual

GX-2012 Operator’s Manual Calibration Mode • 71
calibration initiated because of a failed bump test is the difference
between the bump test gas time and the calibration time defined in the
BUMP--SET menu item in Maintenance Mode.
At the end of the calibration, the instrument displays the results from
both the bump test and the calibration. Use the AIRS or (SHIFT)T
button to scroll between the calibration/bump test results, the bump
test gas readings, and the calibration gas readings.
1
20.9
37
10.0
%LEL
%
ppm
ppm
MAINTENANCE
CH4
O2
CAL 60
CO
H2S
FP
PP
FF
PP
%LEL
%
ppm
ppm
MAINTENANCE
CH4
O2
BUMP/CAL
CO
H2S
1
12.0
5
23.0
%LEL
%
ppm
ppm
MAINTENANCE
CH4
O2
BUMP/
CO
H2S
49
12.0
5
25.0
%LEL
%
ppm
ppm
MAINTENANCE
CH4
O2
/CAL
CO
H2S