Connecting a GX-2009 to the User Setup Program 17
indica te that a wireless connection is in effect.
4. There are two exceptions to the sequence described in
step 3 above. See “Detail Settings” on page 44 for a
description of the Cal Limit Display and Cal Limit Check
parameters.
When Cal Limit Display is set to On and Cal. Limit
Check is set to Can’ t Use, if the GX-20 09 is due for
calibration the instrument will show a calibration expired
failure scr een and the buzzer and LED arrays will pul se for
a few seconds and before connectin g to the program. The
failure screen will remain on the display while the unit is
connected to the program. You can still use the User
Setup Program wi th an instrument if this failure indication
occurs.
When Cal Limit Display is set to On and Cal. Limit
Check is set to Conf irm to use, if the GX-2009 is due for
calibration the unit will show the following screen and beep
and flash the LED arra ys for a few seconds indicati ng tha t
calibration is due.
When the beeping and flashing stops, press and release
the POWER MODE button on the GX-2009 to continue.
The unit will now connect to the program and the PC
TRANSMIT screen will be displayed.
PC
TRANSMIT
FAIL
C-LIMIT
AL