Manual

Control Buttons • 17
The following types of data les are saved in the Data window:
Calibration History Files
A calibration history le is saved for each instrument that has
been downloaded. It records the calibration information for
every calibration that was downloaded. The GX-2003 can
save information for up to 20 calibrations in its memory.
Interval Trend Data Files
Interval trend data is logged at the interval time dened in the
GX-2003. Each logged point is an average reading over the
previous time interval.
Alarm Trend Data Files
Alarm trend data is logged around an alarm event. The GX-
2003 can save up to 8 alarm trend data les in its memory.
Alarm Event Files
Alarm event les record gas alarm events that have been
downloaded from instruments. The GX-2003 can save up to
100 alarm events in its memory.
•Trouble Event Files
Trouble event les record sensor failure and system failure
events that have been downloaded from instruments. The
GX-2003 can save up to 100 trouble events in its memory.