Instruction Manual

Maintainin
g
the GX-2001 Calibration Station 55
Adjusting the Low Flow Alarm
The factory set low ow alarm setting is 0.4 SCFH (±0.1). The
low ow alarm can be adjusted using the low ow adjustment
screw, which is located in the upper left corner of the back panel.
It is accessible through a hole in the panel. Use a small athead
screwdriver to make the adjustment. To adjust the low ow
setting, do the following:
1. Follow the instructions in the Manual Calibration section
above, but do not connect a calibration gas cylinder to the
calibration station.
2. After you press the Cal. Unit Calibration button, as soon as
the Calibration Station begins to drawing fresh air, continue
with step 3.
3. Use the fresh air ow adjustment valve on the back panel of
the Calibration Station to set the ow to 0.4 SCFH.
If the Calibration Station goes into low ow alarm before you
can adjust the ow down to 0.4 SCFH, adjust the low ow
adjustment screw 1/4 turn clockwise, then attempt to set the
ow again. Repeat this step until you are able to adjust the
ow to 0.4 SCFH.
4. Slowly turn the low ow adjustment screw counterclockwise
in very small increments and wait 5 seconds after each
increment to see if the Calibration Station goes into low ow
alarm. Continue this process until the Calibration Station
goes into low ow alarm.
5. Increase the ow using fresh air ow adjustment valve until
the unit is out of low ow alarm.
6. Decrease the ow very slowly and verify that the low ow
alarm is 0.4 SCFH (±0.1).
7. If the low ow alarm is set too high, turn the low ow
adjustment screw slightly clockwise. Repeat steps 5 and 6.
If the low ow alarm is set too low, turn the low ow
adjustment screw slightly counterclockwise. Repeat steps 5
and 6.