Instruction Manual

26 • Down Loadin
g
Data from the Model GX-2001
3. Choose the COM port on the computer to which you have
connected the Data Logging/Charger Station.
4. Press the Mode/Power
button on the Model GX-
2001. “Conn” displays on
the Model GX-2001 LCD
and “GX-2001...Connected
Successful. displays in the
Download area of the
Download window.
5. If you have selected
“Automatic Download, the
downloading process
begins automatically after
you have placed the Model
GX-2001 in the Data
Logging/Charger Station
and after you have turned
on the unit by pressing the
Mode/Power button.
6. If you are going to down load data manually, you can down
load all information at once, or you can load only the
information you want.
•To down load instrument
information only, click
“Instrument Information.
•To down load interval
information, click “Interval
Download.
•To down load alarm trend
data only, click “Alarm
Trend.
Figure 12. Connection
Message
Figure 13. Download
Options