Instruction Manual

Overview of the GX-2001 Calibration Station 19
Computer Connector. The computer receives its information
from the GX-2001 Calibration Station when a DB-9 female/
female extension cable is plugged into this connector and the
other end of the cable is plugged into a serial port on the
computer. A six foot cable is supplied with the Calibration
Station.
115 VAC IN Connector. This connector is for the power cord
that is provided with the Calibration Station.
Fresh Air Flow Adjustment Valve. Using a athead
screwdriver, you can adjust the fresh air ow rate. The
recommended ow rate is 1 SCFH.
Span Gas Flow Adjustment Valve. Using a athead
screwdriver, you can adjust the span gas ow rate. The
recommended ow rate is 1 SCFH.
Low Flow Adjustment Screw. Using a small athead
screwdriver, you can adjust the low ow alarm setting.
NOTE: The low ow alarm setting is factory set and does not
normally need adjustment in the eld. If it does need
adjustment, see the Adjusting the Low Flow Alarm
section for instructions.
Charcoal Filter. When fresh air is drawn into the GX-2001
Calibration Station, it rst passes through the charcoal lter,
which removes traces of gas that might be in your work
environment. This lter is user-replaceable.
Particle Filter. After the fresh air ow passes through the
charcoal lter, it goes through the particle lter, which removes
any dust that might be present in your work area. This lter is
user-replaceable.
Fuse. A 1/4-inch x 1 1/4-inch fast-acting fuse, rated at 1A, 250V,
protects the circuitry of the GX-2001 Calibration Station from
electrical overload. The fuse is user-replaceable.